For testing and calibration laboratories, internal audits are more than a management system requirement. They are one of the main ways a laboratory checks that its results remain valid and that it is ready for assessment by its accreditation body. ISO/IEC 17025:2017, the current edition, requires laboratories to conduct internal audits at planned intervals. This article explains what that involves and how UAE laboratories can make audits effective. For accreditation, see ISO/IEC 17025 accreditation in the UAE.
Accreditation, not certification
ISO/IEC 17025 leads to accreditation, not certification. An accreditation body, such as the Emirates International Accreditation Centre (EIAC) in the UAE, assesses a laboratory's technical competence, impartiality and consistent operation for a defined scope of tests or calibrations. ISO notes that accredited reports and certificates can be accepted between countries without further testing, which is one reason laboratories value accreditation. Internal audits are a key part of maintaining it.
What ISO/IEC 17025 requires
The laboratory must conduct internal audits at planned intervals to provide information on whether its management system conforms to its own requirements, including its laboratory activities, and to the standard, and whether it is effectively implemented and maintained. It must plan an audit programme that takes into account the importance of the activities concerned, changes affecting the laboratory and the results of previous audits; define audit criteria and scope; ensure results are reported to relevant management; implement appropriate correction and corrective action without undue delay; and retain records as evidence.
Two sides of a laboratory audit
Management system
Document control, records, risks and opportunities, improvement, corrective action, complaints, management review and, for laboratories using the second option in the standard, the related ISO 9001-based system.
Technical requirements
Personnel competence and authorisation, facilities and environmental conditions, equipment, metrological traceability, externally provided products and services, review of requests and contracts, selection, verification and validation of methods, sampling, handling of test or calibration items, technical records, evaluation of measurement uncertainty, ensuring the validity of results, reporting, complaints and nonconforming work.
Many laboratories audit the management system thoroughly but give the technical side less attention. Accreditation assessors focus heavily on technical competence, so internal audits should too.
Planning the programme
- Cover every requirement of the standard and every test or calibration area in the scope within the audit cycle.
- Give more attention to complex methods, new methods, areas with past findings and areas affected by changes in staff, equipment or premises.
- Combine horizontal audits, which check one requirement across the laboratory, with vertical audits, which follow one sample or job from request to report.
- Include witnessing of tests or calibrations being performed.
Auditors
Auditors should be trained in ISO/IEC 17025 and audit techniques, and technical audits need people who understand the methods being audited. They must be independent of the activity they audit. Smaller laboratories may need to arrange cross-audits with another laboratory or use an external auditor for some areas.
A vertical audit in practice
- Select a completed report or certificate.
- Check the request, review of the contract and any deviations agreed with the customer.
- Check receipt, identification, storage and handling of the item.
- Check that the method used was the current, verified or validated version.
- Check the analyst's competence and authorisation for the method on that date.
- Check equipment calibration status and intermediate checks.
- Check environmental conditions recorded during the work.
- Check raw data, calculations and measurement uncertainty.
- Check quality control results and proficiency testing for the method.
- Check that the report contains the required information and was reviewed and authorised.
Common findings
- Competence records not covering every method an analyst performs.
- Methods modified without validation.
- Measurement uncertainty estimates outdated or incomplete.
- Equipment intermediate checks not performed between calibrations.
- Environmental condition records missing during critical tests.
- Impartiality risks not reviewed.
- Quality control data recorded but not trended.
- Reports issued without the information the standard requires, such as the method used or deviations.
- Complaints about results handled informally, without records or evaluation of root cause.
Preparing for the accreditation assessment
A well-run internal audit programme is the best preparation for an accreditation assessment or surveillance visit. Before assessors arrive, confirm that the full audit cycle has been completed, that findings are closed with evidence, that management review has considered audit results, and that a sample of vertical audits has been carried out for each method in the scope. Assessors often witness tests and interview analysts, so internal auditors should do the same. Laboratories that treat internal audits as a rehearsal of the external assessment usually find fewer surprises, and they build analysts' confidence in explaining their work.
Acting on results
Correct immediate problems, analyse root causes, act to prevent recurrence and evaluate whether nonconforming work affected results already reported. Where it did, the laboratory must take appropriate action, which may include informing customers. Feed audit results into management review, along with proficiency testing, quality control and complaint data. Sterling engagements come with a free Effivity ISO software subscription for a full year, renewed each year Sterling supports your surveillance or recertification, and free Sterling Academy e-learning for your people, together with complimentary awareness training and internal auditor training as applicable to your standard. Laboratories also seeking ISO 9001 certification can align the two systems; see ISO 9001 certification in the UAE.
Get in touch
For a conversation about internal audits for your laboratory, reach Sterling's Dubai team through the contact page or on +971 50 584 2597. A Sterling consultant replies within one working day.